Table 1: Test Validation Data with the ReaScan Strep A Test Sample


GAS control swab (Bioclin, Finland) GAS control swab (Bioclin, Finland) GAS control swab (Bioclin, Finland)


GAS positive control included in one kit lot*


1 mm x 1 mm GAS colony on blood agar dispensed in 1 ml PBS Extraction reagent blank


n


20 20 20 30 10 10


1 mm x 1 mm Streptococcus group B** colony on blood agar dispensed in 1 ml PBS 10 Negative patient samples***


176 CFU/Swab


1.2 x 105 1.2 x 104 1.2 x 103 N/A N/A 0


N/A 0


ReaScan


Value Range 100–112 40–52 10–19


276–347 220–260 0–3 0–5


0–19 Mean of


ReaScan Value 109 46 13


315 239 1 2 1


* The positive control is prepared to a level corresponding to 1.2 x 105 CFU/swab or more (most commercial RADT tests show positive GAS at this level). ** Similarly, Group C, F and G streptococci, (S. milleri, S. anginosus, S. constellatus, S. intermedius), S. salivarius, Enterococcus faecalis, E. casseliflavus, Staphylococcus aureus and Arcanobacterium haemolyticum gave ReaScan values ≤5.0 (= negative).


*** The samples were confirmed as negative with the standard selective sheep blood agar for streptococci cultivation method. Theoretical sensitivity: 1 x 104 CFU/sample. Specificity: only GAS has been found to be reactive in the test. CFU = colony-forming units; GAS = Group A Streptococcus; N/A = not available; PBS = phosphate-buffered saline.


Figure 1A: Outline of the Test Principle Figure 1B: Outline of the Test Procedure


Gold-conjugated GAS antibody


Dispense three drops of extraction reagents 1 and 2 to the assay vial


GAS antigen in the sample


Immerse the swab into the assay vial and rotate it for one minute


Dispense three drops of extraction reagent 3 into the assay vial


Test membrane GAS = Group A Streptococcus.


by the instrument and therefore the observer’s impaired vision or expectations do not interfere with the results.


ReaScan Strep A


The ReaScan Strep A test (Reagena International Ltd, Toivala, Finland) is a modified immunochromatographic lateral-flow test for the detection of GAS from a regular flocked nylon throat swab (microRheologics, Brescia, Italy). This rapid test detects streptococcal Lancefield group A antigen, which is extracted from the specimen by the extraction reagents included in the test kit. The extraction follows a previously well-described procedure.7


The extracted GAS antigen is allowed to


react with gold particles coated with GAS-specific antibodies. Then other membrane-bound GAS antibodies capture the gold particles with the GAS antigen causing the appearance of a red test line (see Figure 1A). An outline of the test protocol is shown in Figure 1B. The intensity of the test line is measured by a ReaScan reader, which gives a numerical value proportional to the amount of antigen in the sample. Since the test result is based on a numerical value, human errors are minimised, thus increasing the diagnostic value of the ReaScan Strep A test.


The discrepant analysis of the ReaScan Strep A test was performed comparing the rapid test with the standard culture method and


144


Pipette 80 µl from the conjugate vial to the test cassette


Wait 15 minutes and read the result with ReaScan©


reader


with the TestPack® Strep A test (Inverness Medical, Cranfield, UK), according to instructions for use. Three parallel swab samples were taken for each of the study methods from 231 volunteers visiting the outpatient clinic of Oulu University Hospital. The standard culture method8


Pipette 100 µl from the assay vial to the conjugate vial and mix well


results were classified as negative or positive (+, ++ or +++). In this method, sheep blood agar for streptococci is used (5 % sheep blood agar, Müller–Hinton + tryptone–soya agar), supplemented with colistin and nalidixic acid (IS-ST-A, Abtek Biological Ltd, UK) and with supplemental carbon dioxide to enhance the growth of streptococci. Streptococcal Grouping Reagent A (DR0586G, Oxoid Ltd, UK) was used.


Results and Discussion


In the present study, we introduced a novel immunochromatographic bacterial Strep A antigen test for the ReaScan rapid test reader. We evaluated the sensitivity, specificity and cross-reactivity of the novel rapid test. Table 1 summarises the test validation data with the ReaScan Strep A test. No cross-reactions with cultured Group B,


EUROPEAN INFECTIOUS DISEASE


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92